7.4 Extended Event Register
7-8
The transition filter parameters detect changes in the specified bit (numerical suffix, 1 to
16) of the condition register in the following manner and overwrite the extended event
register.
RISE
The specified bit of the extended event register is set to 1 when the bit of the
condition register changes from 0 to 1.
FALL
The specified bit of the extended event register is set to 1 when the bit of the
condition register changes from 1 to 0.
BOTH
The specified bit of the extended event register is set to 1 when the bit of the
condition register changes from 0 to 1 or 1 to 0.
NEVer
Always 0.
IM WT3001E-17EN