YOKOGAWA WT3001E User Manual page 100

Precision power analyzer communication interface
Hide thumbs Also See for WT3001E:
Table of Contents

Advertisement

6.9 FLICker Group
:FLICker:STARt
Function
Starts flicker measurement.
:FLICker:STARt
Syntax
:FLICKER:START
Example
:FLICker:STATe?
Function
Queries the status of flicker measurement.
:FLICker:STATe?
Syntax
:FLICKER:STATE? -> RESET
Example
Description The contents of the response are as follows:
RESet = Reset status
INITialize = Initializing
READy = Measurement start wait state
STARt = Measuring
COMPlete = Measurement stopped,
:FLICker:TMAX?
Function
Queries all settings related to Tmax.
:FLICker:TMAX?
Syntax
:FLICKER:TMAX? ->
Example
:FLICKER:TMAX:STATE 1;LIMIT 500,3.30
Description This is the same query as with the
" :FLICker:DT? " command.
:FLICker:TMAX:LIMit
Function
Sets the limit of the Tmax or queries the current
setting.
:FLICker:TMAX:LIMit {<NRf>:LIMit
Syntax
{<NRf>[,<NRf>]}
:FLICker:TMAX:LIMit?
1st <NRf> = 1 to 99999 (limit [ms])
2nd <NRf> = 1.00 to 99.99 (threshold level [%])
:FLICKER:TMAX:LIMIT 500,3.30
Example
:FLICKER:TMAX:LIMIT? ->
:FLICKER:TMAX:LIMIT 500,3.30
Description This is the same setting or query as with the
" :FLICker:DT:LIMit " command.
:FLICker:TMAX[:STATe]
Function
Turns ON/OFF judgment of the Tmax or queries
the current setting.
:FLICker:TMAX[:STATe] {<Boolean>}
Syntax
:FLICker:TMAX:STATe?
:FLICKER:TMAX:STATE ON
Example
:FLICKER:TMAX:STATE? ->
:FLICKER:TMAX:STATE 1
Description This is the same setting or query as with the
" :FLICker:DT[:STATe] " command.
:FLICker:UN?
Function
Queries all settings related to rated voltage Un.
:FLICker:UN?
Syntax
:FLICKER:UN? -> :FLICKER:UN:
Example
MODE AUTO;VALUE 230.00
6-56
judgment results displayed
:FLICker:UN:MODE
Function
Sets the assignment method for rated voltage Un
or queries the current setting.
:FLICker:UN:MODE {AUTO|SET}
Syntax
:FLICker:UN:MODE?
AUTO = Use the voltage value measured upon
start of measurement.
SET = Use the predefined value
(:FLICker:UN:VALue).
:FLICKER:UN:MODE AUTO
Example
:FLICKER:UN:MODE? ->
:FLICKER:UN:MODE AUTO
:FLICker:UN:VALue
Function
Sets the predefined value of rated voltage Un or
queries the current setting.
:FLICker:UN:VALue {<NRf>}
Syntax
:FLICker:UN:VALue?
<NRf> = 0.01 to 999.99 (predefined value[V])
:FLICKER:UN:VALUE 230.00
Example
:FLICKER:UN:VALUE? ->
:FLICKER:UN:VALUE 230.00
:FLICker:VOLTage
Function
Sets the flicker target voltage or queries the
current setting.
:FLICker:VOLTage {<NRf>}
Syntax
:FLICker:VOLTage?
<NRf> = 120, 230 (target voltage[V])
:FLICKER:VOLTAGE 230
Example
:FLICKER:VOLTAGE? ->
:FLICKER:VOLTAGE 230
IM WT3001E-17EN