YOKOGAWA WT3001E User Manual page 139

Precision power analyzer communication interface
Hide thumbs Also See for WT3001E:
Table of Contents

Advertisement

:NUMeric:NORMal?
Function
Queries all settings related to the numeric data
output.
:NUMeric:NORMal?
Syntax
:NUMERIC:NORMAL? -> :NUMERIC:
Example
NORMAL:NUMBER 15;ITEM1 U,1,TOTAL;
ITEM2 I,1,TOTAL;ITEM3 P,1,TOTAL;
ITEM4 S,1,TOTAL;ITEM5 Q,1,TOTAL;
ITEM6 LAMBDA,1,TOTAL;
ITEM7 PHI,1,TOTAL;ITEM8 FU,1;
ITEM9 FI,1;ITEM10 UPPEAK,1;
ITEM11 UMPEAK,1;ITEM12 IPPEAK,1;
ITEM13 IMPEAK,1;ITEM14 CFU,1;
ITEM15 CFI,1
Description For the values of
" :NUMeric[:NORMal]:ITEM<x>, " the numeric
data output items for the amount specified by
" :NUMeric[:NORMal]:NUMber " are output.
:NUMeric[:NORMal]:CLEar
Function
Clears the numeric data output item (sets
"NONE").
:NUMeric[:NORMal]:CLEar {ALL|
Syntax
<NRf>[,<NRf>]}
ALL = Clear all items
1st <NRf> = 1 to 255 (Item number to start
clearing)
2nd <NRf> = 1 to 255 (Item number to end
clearing)
:NUMERIC:NORMAL:CLEAR ALL
Example
Description If the 2nd <NRf> is omitted, the output items from
the start clear number to the last item (255) are
cleared.
:NUMeric[:NORMal]:DELete
Function
Deletes the output items of numeric data.
:NUMeric[:NORMal]:DELete {<NRf>
Syntax
[,<NRf>]}
1st <NRf> = 1 to 255 (Item number to start
deleting)
2nd <NRf> = 1 to 255 (Item number to end
deleting)
:NUMERIC:NORMAL:CLEAR 1 (Deletes
Example
ITEM1 and shift ITEM2 and subsequent
items forward)
:NUMERIC:NORMAL:CLEAR 1,3 (Deletes
ITEM1 to ITEM3 and shift ITEM4 and subsequent
items forward)
Description • The subsequent output items fill the positions
of deleted output items, and empty sections at
the end are set to "NONE."
• If the 2nd <NRf> is omitted, only the output
item of the delete start number is deleted.
IM WT3001E-17EN
:NUMeric[:NORMal]:ITEM<x>
Function
Sets the numeric data output items (function,
element, and harmonic order) or queries the
current setting.
:NUMeric[:NORMal]:ITEM<x> {NONE|
Syntax
<Function>,<Element>[,<Order>]}
:NUMeric[:NORMal]:ITEM<x>?
<x> = 1 to 255 (item number)
NONE = No output item
<Function> = {U|I|P|S|Q|...} (See the
function selection list (1) of "DISPlay group" on
page 6-44.)
<Element> = {<NRf>|SIGMa|SIGMB} (<NRf>
= 1 to 4)
<Order> = {TOTal|DC|<NRf>} (<NRf> = 1 to
100)
:NUMERIC:NORMAL:ITEM1 U,1,TOTAL
Example
:NUMERIC:NORMAL:ITEM1? ->
:NUMERIC:NORMAL:ITEM1 U,1,TOTAL
Description • If <Element> is omitted, element 1 is set.
• If <Order> is omitted, TOTal is set.
• <Element> or <Order> is omitted from
response to functions that do not need them.
:NUMeric[:NORMal]:NUMber
Function
Sets the number of the numeric data that is
transmitted by ":NUMeric[:NORMal]:VALue?" or
queries the current setting.
:NUMeric[:NORMal]:NUMber {<NRf>|ALL}
Syntax
:NUMeric[:NORMal]:NUMber?
<NRf> = 1 to 255(ALL)
:NUMERIC:NORMAL:NUMBER 15
Example
:NUMERIC:NORMAL:NUMBER ->
:NUMERIC:NORMAL:NUMBER 15
Description • If the parameter is omitted for the
" :NUMeric[:NORMal]:VALue? " command,
the numeric data from 1 to (the specified value)
is output in order.
• By default, the number of numeric data is set to
"15."
:NUMeric[:NORMal]:PRESet
Function
Presets the output item pattern of numeric data.
:NUMeric[:NORMal]:PRESet {<NRf>}
Syntax
<NRf> = 1 to 4
:NUMERIC:NORMAL:PRESET 1
Example
Description • For details on the output items that are preset,
see "(1) Preset Pattern of Numeric Data Output
Items" (page 6-98).
• By default, output items of "Pattern 2" is
selected.
6.18 NUMeric Group
App
Index
6-95
1
2
3
4
5
6
7

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wt3002eWt3003eWt3004e

Table of Contents