:NUMeric:LIST:DELete
Function
Deletes the output items of the numeric list data
of harmonic measurement.
:NUMeric:LIST:DELete {<NRf>
Syntax
[,<NRf>]}
1st <NRf> = 1 to 64 (Item number to start
deleting)
2nd <NRf> = 1 to 64 (Item number to end
deleting)
:NUMERIC:LIST:CLEAR 1 (Deletes ITEM1
Example
and shift ITEM2 and subsequent items
forward)
:NUMERIC:LIST:CLEAR 1,3 (Deletes ITEM1
to ITEM3 and shift ITEM4 and subsequent items
forward)
Description • This command is valid only on models with the
advanced computation function (/G6 option).
• The subsequent output items fill the positions
of deleted output items, and empty sections at
the end are set to "NONE."
• If the 2nd <NRf> is omitted, only the output
item of the delete start number is deleted.
:NUMeric:LIST:ITEM<x>
Function
Sets the output items (function elements) of the
numeric list data of harmonic measurement or
queries the current setting.
:NUMeric:LIST:ITEM<x> {NONE|
Syntax
<Function>,<Element>}
:NUMeric:LIST:ITEM<x>?
<x> = 1 to 64 (item number)
NONE = No output item
<Function> = {U|I|P|S|Q|LAMBda|PHI|
PHIU|PHII|Z|RS|XS|RP|XP|UHDF|IHDF|
PHDF}
<Element> = {<NRf>|SIGMa|SIGMB} (<NRf>
= 1 to 4)
:NUMERIC:LIST:ITEM1 U,1
Example
:NUMERIC:LIST:ITEM1? ->
:NUMERIC:LIST:ITEM1 U,1
Description This command is valid only on models with the
advanced computation function (/G6 option).
IM WT3001E-17EN
:NUMeric:LIST:NUMber
Function
Sets the number of the numeric list data that
is transmitted by ":NUMeric:LIST:VALue?" or
queries the current setting.
:NUMeric:LIST:NUMber {<NRf>|ALL}
Syntax
:NUMeric:LIST:NUMber?
<NRf> = 1 to 64 (ALL)
:NUMERIC:LIST:NUMBER 5
Example
:NUMERIC:LIST:NUMBER ->
:NUMERIC:LIST:NUMBER 5
Description • This command is valid only on models with the
advanced computation function (/G6 option).
• If the parameter is omitted for the
" :NUMeric:LIST:VALue? " command, the
numeric list data from 1 to (the specified value)
is output in order.
• By default, the number of numeric data is set to
"1."
:NUMeric:LIST:ORDer
Function
Sets the maximum output order of the numeric
list data of harmonic measurement or queries the
current setting.
:NUMeric:LIST:ORDer {<NRf>|ALL}
Syntax
:NUMeric:LIST:ORDer?
<NRf> = 1 to 100(ALL)
:NUMERIC:LIST:ORDER 100
Example
:NUMERIC:LIST:ORDER? ->
:NUMERIC:LIST:ORDER 100
Description This command is valid only on models with the
advanced computation function (/G6 option).
:NUMeric:LIST:PRESet
Function
Sets the output items of harmonic measurement
numeric list data to a preset pattern.
:NUMeric:LIST:PRESet {<NRf>}
Syntax
<NRf> = 1 to 4
:NUMERIC:LIST:PRESET 1
Example
Description • This command is valid only on models with the
advanced computation function (/G6 option).
• For details on the output items that are preset,
see "(2) Preset Pattern of the Numeric List
Data Output Items of Harmonic Measurement "
(see page 6-100).
• By default, output items of "Pattern 2" is
selected.