• The contents of the selections for <Element>
are as follows. When <Element> is omitted,
Element 1 is set.
TOTal The overall judgment result for all
measured elements is only available
when <Function> = JTOTal (judgment
result displayed under [Total Judgment:]
in the upper right part of the screen)
<NRf> = 1 to 4 specified elements
• The contents of the selections for <Period> are
as follows.
If <Period> is omitted, ALL is set.
ALL Overall observation period (Result row
of the numeric list in the flicker measurement
display screen)
<NRf> = 1 to 99 specified observation periods
• <Period> is omitted from the response to
output items in the table above for which
specification of <Period> is not required.
• JTMAX, a new function defined in IEC61000-3-
3 Ed3.0, can be used. JTMAX represents the
same content as the conventional DT function,
and the judgment results that is output is also
the same.
• For IEC61000-3-3 Ed3.0, "JTMAX" is returned
as a response to a setting query.
:NUMeric:FLICker:INFOrmation:NUMber
Function
Sets the number of flicker judgment results output
by ":NUMeric:FLICker:INFOrmation:VALue?" or
queries the current setting.
:NUMeric:FLICker:INFOrmation:
Syntax
NUMber {<NRf>|ALL}
:NUMeric:FLICker:INFOrmation:
NUMber?
<NRf> = 1 to 32(ALL)
:NUMERIC:FLICKER:INFORMATION:
Example
NUMBER 7
:NUMERIC:FLICKER:INFORMATION:
NUMBER -> :NUMERIC:FLICKER:
FUNCTION:NUMBER 7
Description • Only available with the flicker measurement
function (/FL option).
• If parameters are omitted from " :NUMeric:
FLICker:INFOrmation:VALue? ", judgment
results from 1 to (specified value) are output in
order.
• The initial setting for the number of judgment
results is 7.
IM WT3001E-17EN
:NUMeric:FLICker:INFOrmation:VALue?
Function
Queries the judgment results (variable format).
:NUMeric:FLICker:INFOrmation:
Syntax
VALue? {<NRf>}
:NUMeric:FLICker:INFOrmation:VALue?
<NRf> = 1 to 32 (item number)
Example
• Example when <NRf> is specified
:NUMERIC:FLICKER:INFORMATION:
VALUE? 1 -> 0
• Example when <NRf> is omitted
:NUMERIC:FLICKER:INFORMATION:
VALUE? -> 0,0,0,0,0,0,0
• Example in which " :NUMeric:FORMat " is set
to " FLOat ".
:NUMERIC:FLICKER:INFORMATION:
VALUE? -> #4 (number of bytes, four digits)
(data byte sequence)
Description • Only available with the flicker measurement
function (/FL option).
• When <NRf> is specified, only the judgment
results for that item is output.
• If <NRf> is omitted, judgment results from the
item number in
" :NUMeric:FLICker:INFOrmation:
NUMber " is output in order.
• The format of individual judgment results that
are output is as follows:
• Judgment result (JTOTal, JDC, JDMAX,
JDT, JPST, JPLT)
ASCII
FLOAT
(<NR1> format)
(IEEE single-precision floating
point (4-byte) format)
Pass:
"0"
0x00000000 (0)
Fail:
"-1"
0xBF800000 (-1)
Error:
"-2"
0xC0000000 (-2)
-----:
"1"
0x3F800000 (1)
(space):
"1"
0x3F800000 (1)
• No items (NONE)
ASCII: "NAN" (Not A Number)
FLOAT: 0x7E951BEE (9.91E+37)