PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B

Network Analyzer System Operation

The PNA series network analyzer generates a phase-locked incident signal from the
internal synthesized source. By means of signal separation, the incident signal is divided
into a reference signal and a test signal. The reference signal is applied to the receiver
group, while the test signal is applied to the device under test (DUT) and then to the
receiver group. The signals are downconverted and are next sampled and digitally
processed.
Figure 5-1
Figure 5-1 System Simplified Block Diagram

Functional Groups of the Network Analyzer

The operation of the network analyzer can be separated into six functional groups. Each
group consists of several major assemblies that perform a distinct function in the
instrument. Some of the assemblies are related to more than one group, and all the groups,
to some extent, are interrelated and affect each other's performance. The six major
functional groups are:
Synthesized Source Group
Signal Separation Group
Receiver Group
Frequency Offset Group
Digital Processor and Digital Control Group
Power Supply Group
Service Guide E8364-90026
is a simplified block diagram of the network analyzer system.
Theory of Operation
Network Analyzer System Operation