Solution
Restart the Cary 60. The calibration procedure occurs automatically on start up.
Agilent recommends that as part of your standard operating procedure you restart
the Cary 60 monthly. Perform the Instrument Performance Test suite in 'Validate'
to ensure the instrument results meet specification.
Front Panel LED Troubleshooting
LED Won't Light Up
No power to the instrument.
Check power cables and power point switches (if applicable).
Solid Red LED
The instrument is available to use but the previous initialization or calibration has
failed.
Initialization or calibration failed.
Check or perform the following:
The sample compartment for items blocking the beam.
•
The cuvette is inserted correctly and clean.
•
Reboot the Cary 60.
•
Contact your local Agilent field service engineer or representative.
•
Flashing Red LED
The instrument is scanning but the previous initialization or calibration has failed.
Initialization or calibration failed.
Check or perform the following:
The sample compartment for items blocking the beam.
•
The cuvette is inserted correctly and clean.
•
Reboot the Cary 60.
•
Contact your local Agilent field service engineer or representative.
•