Monitoring Of Additional Breathing Parameters - Siare Aria 104 User Manual

Lung ventilator
Table of Contents

Advertisement

2.3.8

Monitoring of Additional breathing parameters

2-26
Based on the lung ventilator parameters set by the User and on the
patient's characteristics, the lung ventilator is able to monitor and
measure a series of values necessary for the patient's clinical
evaluation.
The measured and monitored values (right side of the screen) are
updated after each breath of the patient.
E.g. Here below the data in the images below refer to VC/VAC operative
mode.
Static compliance
It is one of the parameters of the lung mechanics: measured in
ml/cmH
O, represents the lung compliance when the patient does
2
not breathe.
You can use it to asses the lung elasticity: the higher the
compliance, the more elastic the "lung"; the lower the compliance,
the more "rigid" the lung.
The static compliance can be calculated using the formula below:
Cs = current inspired volume / pause pressure.
Dynamic compliance
It is one of the parameters of the lung mechanics: measured in
ml/cmH
O, represents the lung compliance either during the
2
inspiration or during the expiration.
You can use it to asses the lung elasticity: the higher the
compliance, the more elastic the "lung"; the lower the compliance,
the more "rigid" the lung. The dynamic compliance can be
calculated using the formula below:
Cd = current inspired volume / peak pressure
User manual, DU3104101