Periodic Maintenance; Maintenance Operations - Siare Falco 202 Evo User Manual

Lung ventilator
Table of Contents

Advertisement

6.4

Periodic maintenance

Inspections and periodic maintenance are ensured by taking out a maintenance
contract with SIARE or an authorised dealer.
Contact SIARE for information regarding authorised Service Centres in your area.
When you require service, please indicate the serial number of the unit and the problem
to SIARE or to your authorised technicians.
SIARE assumes responsibility for all provisions foreseen by the law, if the equipment is
used and maintained as per the instructions in this manual and the technical manual
The Technical Assistance Report, signed by the authorised SIARE technician, is proof
of the completion of the scheduled maintenance.

6.4.1 Maintenance operations

The table summarizes the preventive maintenance frequency and procedures to be carried
out on the lung ventilator.
Frequency
Several times a day
/ according to local
practice and
standards
6-10
CAUTION
The lung ventilator does not require particular maintenance and preventive
operations other than those indicated in this manual or in order to respect
standards applied in the specific country where the lung ventilator is sold.
WARNING !!
Risk of injury for the patient
Always refer to the instructions contained in the previous section: cleaning,
disinfection and sterilization of the components.
Component
Patient circuit
Disposable bacteria
filter
Condensation trap
filter
Procedure / Action
Check for any water collection, drain and clean the
tubes when necessary.
Replace.
Check for any water collection, drain and clean
when necessary.