Qiagen QIAcuity 911000 User Manual page 77

Hide thumbs Also See for QIAcuity 911000:
Table of Contents

Advertisement

6a.
In the Imaging tab, select the applicable channel, then enter the exposure duration and
gain in the Exposure duration and Gain fields.
6b.
To include more steps in the run, tap the Add
Provide the required information for the step. Repeat this step if more steps are needed
for the run. In total 9 steps can be performed per plate.
6c.
Tap Save to save your progress or tap Done to save the run and go back to the Running
Status window.
Note: If any required field is not completed, an error message displays to list the missing
information required in each field.
7. Start the run in the Running Status window:
To start the run on all plates simultaneously without making any changes, tap Run all.
To start the run of an individual plate without making any changes, tap the corresponding
Run
icon located on the plate's pane.
Linking a plate to a pre-defined plate without existing barcode
If the instrument cannot match the barcode on a loaded plate to a barcode that already exists in
the Software Suite, you can link the plate manually. Alternatively, you can create a new plate by
following the steps in section Configuring a plate and starting a run.
QIAcuity User Manual 07/2020