Scheme of filling and partitioning of a well.
Principle of priming and rolling to allow partitioning of the wells.
Thermocycling: The second step and module is a high accuracy plate thermocycler that performs
the polymerase chain reaction. The cycling profile can be set in the QIAcuity Software Suite or the
instrument software. For more details on the thermal cycler specification, see Appendix A –
Technical Data
Imaging: The final process step is the image acquisition of all wells. The user can select the detection
channels in the experiment setup. The partitions that have a target molecule inside, emit fluorescence
light and are brighter than the ones without target (see image below). For more details and
specifications on the imaging system, see section Appendix A – Technical Data.
QIAcuity User Manual 07/2020
18