WARNING
Damage to electronics
Before powering ON the instrument, make sure that the correct
supply voltage is used.
Use of incorrect supply voltage may damage the electronics.
To check the recommended supply voltage, refer to the
specifications indicated in the type plate of the instrument.
WARNING
Electrical hazard
Any interruption of the protective conductor (earth/ground lead)
inside or outside the instrument or disconnection of the protective
conductor terminal is likely to make the instrument dangerous.
Intentional interruption is prohibited.
Lethal voltages inside the instrument
When the instrument is connected to line power, terminals may be
live and opening covers or removing parts is likely to expose live
parts.
Grounding requirements
To protect operating personnel, the National Electrical Manufacturers' Association (NEMA)
recommends that the QIAcuity be correctly grounded (earthed). The instrument is equipped with a
3-conductor AC power cord that, when connected to an appropriate AC power outlet, grounds
(earths) the instrument. To preserve this protection feature, do not operate the instrument from an
AC power outlet that has no ground (earth) connection.
QIAcuity User Manual 07/2020
[W9]