Back of QIAcuity
Connecting the power cord to the back of the QIAcuity
1. Remove the power cord from the foam packing material on top of the QIAcuity.
Note: Only use the power cord that is supplied with the QIAcuity.
2. Check that the voltage rating on the label found at the back of the QIAcuity matches the
voltage available at the installation site.
3. Connect the power cord to the power inlet of the instrument and connect the cable to the wall
power outlet.
4. Turn on the power switch at the back of the instrument.
WARNING
Damage to electronics
Before powering ON the instrument, make sure that the correct
supply voltage is used.
Use of incorrect supply voltage may damage the electronics.
To check the recommended supply voltage, refer to the
specifications indicated in the type plate of the instrument.
QIAcuity User Manual 07/2020
[W9]
30