Error Handling

268
Events and errors within a instrument control program can be detected by
polling (querying) the instrument. Polling is used in application
development environments (ADEs) that do not support asynchronous
activities where callbacks can be used.
Programs can set up and use polling as shown below.
1
Declare a variable to contain the function completion code.
ViStatus errStatus;
Every driver function returns the completion code ViStatus.
If the function executes with no I/O errors, driver errors, or instrument
errors, ViStatus is 0 (VI_SUCCESS).
If an error occurs, ViStatus is a negative error code.
Warnings are positive error codes, and indicate the operation
succeeded but special conditions exist.
2
Enable automatic instrument error checking following each function
call.
hp816x_errorQueryDetect
(instrumentHandle, VI_TRUE);
When enabled, the driver queries the instrument for an error condition
before returning from the function.
If an error occurred, errStatus (Step Figure 1) will contain a code
indicating that an error was detected
(hp816x_INSTR_ERROR_DETECTED).
3
Check for an error (or event) after each function.
errStatus = hp816x_cmd(instrumentHandle, "SENS1:POW:RANG");
check(instrumentHandle, errStatus);
After the function executes, errStatus contains the completion code.
The completion code and instrument ID are passed to an error
checking routine. In the above statement, the routine is called 'check'.
4
Create a routine to respond to the error or event. This example queries
whether an error has occured, checks if the error is an instrument error
and then checks if the error is a driver error.
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide