Keysight Technologies 8163A/B Programming Manual page 180

Table of Contents

Advertisement

4
Measurement Operations & Settings
command:
:SOURce<SlotNr>:TRANsmitter:RECalibration?
syntax:
:SOURce<SlotNr>:TRANsmitter:RECalibration?
description:
Recalibrates a reference transmitter module.
Recalibration takes about 6.8 seconds, so it also takes this time to get the reply to this query.
parameters:
none
response:
The result is returned as an unquoted string:
"OK" or "ERROR - <error message>".
example:
:SOUR1:TRAN:REC?
affects:
81490A
command:
:SOURce<SlotNr>:TRANsmitter:OPOint<laser (1,2)>
syntax:
:SOURce<SlotNr>:TRANsmitter:OPOint<laser (1,2)> <data>
description:
Sets the operating point for one of the reference transmitter's two laser diodes.
You can also set the minimum, maximum or default value.
parameters:
Operating point or MINimum | MAXimum | DEFault.
response:
none
example:
:sour1:tran:opo1 -17
affects:
81490A
command:
:SOURce<SlotNr>:TRANsmitter:OPOint<laser (1,2)>?
syntax:
:SOURce<SlotNr>:TRANsmitter:OPOint<laser (1,2)>?
description:
Get the operating point for one of the reference transmitter's two laser diodes.
You can also query the minimum, maximum or default value.
parameters:
none | MINimum | MAXimum | DEFault
response:
The operating point.
example:
:sour1:tran:opo2
affects:
81490A