4
Measurement Operations & Settings
command:
:SENSe[n]:[CHANnel[m]]:POWer:RANGe[:UPPer]?
syntax:
:SENSe[n]:[CHANnel[m]]:POWer:RANGe[:UPPer]?
description:
Returns the range setting for the module. For a return loss module, returns the power range of the return loss diode.
parameters:
none
response:
The range setting as a float value in dBm
(-110
example:
sens1:pow:rang?
affects:
All power meters and return loss modules.
dual sensors:
Master and slave channels are independent.
command:
:SENSe[n]:[CHANnel[m]]:POWer:RANGe:MONitor[:UPPer]
syntax:
:SENSe[n]:[CHANnel[m]]:POWer:RANGe:MONitor[:UPPer]<wsp><value>[DBM]
description:
Sets the power range for a retun loss module's monitor diode.
The range changes at 10 dBm intervals. The corresponding ranges for linear measurements (measurements in Watts) is
given below:
Range
+30 dBm
+20 dBm
+10 dBm
0 dBm
-10 dBm
-20 dBm
-30 dBm
-40 dBm
parameters:
The range as a float value in dBm. The number is rounded to the closest multiple of 10, because the range changes at
10 dBm intervals. Units are in dBm.
response:
none
example:
sens1:pow:rang:mon -20DBM
affects:
All return loss modules.
dual sensors:
Master and slave channels are independent.
108
≤
≤
value
+30).
->
-2.00000000E+001<END>
Upper Linear
Power Limit
1999.9 mW
199.99 mW
19.999 mW
m
1999.9
W
m
199.99
W
m
19.999
W
1999.9 nW
199.99 nW
Range
-50 dBm
-60 dBm
-70 dBm
-80 dBm
-90 dBm
-100 dBm
-110 dBm
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
Upper Linear
Power Limit
19.999 nW
1999.9 pW
199.99 pW
19.999 pW
1.999 pW
0.199 pW
0.019 pW