Keysight Technologies 8163A/B Programming Manual page 148

Table of Contents

Advertisement

4
Measurement Operations & Settings
command:
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]:DARK?
syntax:
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]:DARK?
description:
Queries whether the attenuator is set to 'dark' position (where all light is blocked by the laser).
parameters:
none
response:
0:
1:
example:
sour0:pow:att:dark? -> 1<END>
affects:
All tunable laser modules with a built-in optical attenuator.
command:
[:SOURce[n]][:CHANnel[m]]:POWer[:LEVel][:IMMed iate][:AMPLitude[l]]
syntax:
[:SOURce[n]][:CHANnel[m]]:POWer[:LEVel][:IMMediate][:AMPLitude[l]]<wsp><value>
[PW|NW|UW|MW|Watt|DBM]
description:
Sets the power of the laser output.
If an optical attenuator is installed, the power value returned is dependent on whether you are using power or attenuation
mode (see
If you are using power mode, the value returned is the output power.
If you are using attenuation mode, the value returned is the laser output power, and you must also use the attenuation value
to calculate the output power (see
The values for the output power that you set in the Power Mode, and the laser output power that you set in the Attenuation
Mode, are stored and used independently.
The instrument may not be able to output a signal with the maximum programmable power, it will output a signal with the
maximum power. Use the
on page 149 to query the power being output.
The default units are DBM or W, depending on the unit selected using the following command:
[:SOURce[n]][:CHANnel[m]]:POWer:UNIT
parameters:
Any value in the specified range (see the appropriate User's Guide).
Also allowed are:
Use [l] to set the amplitude level of the output power of the upper or lower wavelength laser source of a dual-wavelength
laser source or a return loss module with an internal dual-wavelength laser source. The default value of [l] is 1, the lower
wavelength source. The upper wavelength source is denoted by 2.
response:
none
example:
sour2:pow 23uW
affects:
All tunable laser including 81950A and DFB source modules
148
dark position not set
dark position set
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]:AUTO
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]
[:SOURce[n]][:CHANnel[m]]:POWer[:LEVel][:IMMediate][:AMPLitude[l]]?
MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value, but is the maximum programmable level
on page 151.
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
on page 146).
on page 145).

Advertisement

Table of Contents
loading

This manual is also suitable for:

8164a/b8166a/b

Table of Contents