Keysight Technologies 8163A/B Programming Manual page 51

Table of Contents

Advertisement

command:
*OPC?
syntax:
*OPC?
description:
The OPeration Complete query *OPC? parses all program message units in the input queue, sets the operation complete bit
in the Standard Event Status register, and places an ASCII '1' in the output queue, when the contents of the input queue
have been processed.
Some module firmware includes commands that set a "StatNOPC" flag during execution to indicate that the module is
busy. If a module is executing a command which generates a "StatNOPC" flag, the GPIB bus is not blocked to a command to
another module. A second command to a busy module is blocked until the module flag "StatOK" is set. Taking ad vantage of
this feature, and using *OPC? in a loop to query until the instrument returns 1, can lead to useful gains in program execution
efficiency.
The following actions cancel the *OPC? query (and put the instrument into Operation Complete, Command Idle State):
Power-on
the Device Clear Active State is asserted on the interface.
*CLS
*RST
parameters:
none
response:
1<END> is returned if all modules are ready to execute a new operation.
0<END> is returned if any module is busy.
example:
*OPC?
command:
*OPT?
syntax:
*OPT?
description:
The OPTions query *OPT? returns the modules installed in your instrument.
parameters:
none
response:
Returns the part number of all installed modules, separated by commas.
Slots are listed starting with the lowest slot number, that is, slot 0 for the 8164A/B and Slot 1 for the 8163A/B and 8166A/B.
If any slot is empty or not recognised, two spaces are inserted instead of the module's part number. See the example below,
where slots 1 and 4 are empty.
example:
*OPT?
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
->
1<END>
->
81682A , , 81533B, 81532A, <END>
Instrument Setup and Status
3
51