1
0
example:
stat:ques:level1?
command:
:STATus:QUEStionable:CONDition:LEVel1?
syntax:
:STATus:QUEStionable:CONDition:LEVel1?
description:
Returns the Questionable Status Condition Summary Register for slots 15 to 17 of the 8166A/B Lightwave Multichannel
System.
parameters:
none
response:
The sum of the results for the slots (a 16-bit signed integer value, where 0
Bits
15-4
3
2
1
0
example:
stat:ques:cond:level1?
command:
:STATus:QUEStionable:ENABle:LEVel1
syntax:
:STATus:QUEStionable:ENABle:LEVel1<wsp><value>
description:
Sets the bits in the Questionable Status Enable Summary Mask (QSESM) that enable the contents of the QSESR for slots 15
- 17 of the 8166A/B Lightwave Measurement System to affect the Status Byte (STB).
Setting a bit in this register to 1 enables the corresponding bit in the OSESR for slots 15 - 17 of the 8166A/B Lightwave
Measurement System to affect bit 7 of the Status Byte.
parameters:
The bit value for the QSESM as a 16-bit signed integer value (0 .. +32767)
The default value is 0.
response:
none
example:
stat:oper:enab:level1 128
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
Slot 15 Summary
Not used
->
+0<END>
Mnemonics
8166A/B
Not used
Slot 17 Summary
Slot 16 Summary
Slot 15 Summary
Not used
->
+0<END>
2
0
Decimal Value
0
8
4
2
0
Instrument Setup and Status
value
32767):
3
65