4
Measurement Operations & Settings

Triggering - The TRIGger Subsystem

Hard ware
Trigger Rearming
Triggering
trig:inp
trig:inp:rearm
IGNore
-
SMEasure
ON
CMEasure
ON
SMEasure
OFF
CMEasure
OFF
208
The TRIGger Subsystem allows you to configure how the instrument reacts
to incoming or outgoing triggers.
Table 8
Triggering and Power Measurements
Software Triggering
init:imm
init:cont
One power
Automatically performs power
measurement is
measurements.
performed.
Every hardware trigger starts a new power measurement.
The first hard ware trigger starts a new power measurement.
Further hardware triggers are ignored until you send
trig:inp:rearm again.
Data Acquisition Functions
sens:func:stat
MINMax
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
LOGGing|STABility
Automatically performs power
measurements until the function is
finished.
Every hardware trigger starts a new
power measurement until the
function is finished.
The first hard ware trigger starts the
function. Subsequent power
measurements are automatically
performed until the function is
finished.
Every hardware trigger starts a new
power measurement until the
function is finished.
The first hard ware trigger starts the
function. Subsequent power
measurements are automatically
performed until the function is
finished.