parameters:
The bit value for the OSESM as a 16-bit signed integer value (0 .. +32767)
The default value is 0.
response:
none
example:
stat:oper:enab:level1 128
command:
:STATus:OPERation:ENABle:LEVel1?
syntax:
:STATus:OPERation:ENABle:LEVel1?
description:
Returns the OSESM for the OSESR for slots 15 - 17 of the 8166A/B Lightwave Measurement System
parameters:
none
response:
The bit value for the operation enable mask as a 16-bit signed integer value (0 .. +32767)
example:
stat:oper:enab:level1?
command:
:STATusn:OPERation[:EVENt]?
syntax:
:STATusn:OPERation[:EVENt]?
description:
Returns the Operational Slot Status Event Register (OSSER) of slot n.
parameters:
none
response:
The results for the individual slot events (a 16-bit signed integer value, where 0
Bit
8-15
7
6
5
4
3
2
1
0
example:
stat1:oper?
command:
:STATusn:OPERation:CONDition?
syntax:
:STATusn:OPERation:CONDition?
description:
Returns the Operational Slot Status Condition Register of slot n.
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
->
+128<END>
Mnemonic
Not used
Slot n: offset (λ) type bit 2
Slot n: offset (λ) type bit 1
Slot n: offset (λ) has been enabled
Slot n: shutter has been opened
Slot n: Zeroing ongoing
Not used
Slot n: Coherence Control has been switched on
Slot n: Laser has been switched on
->
+0<END>
Instrument Setup and Status
£
£
value
32767):
Decimal Value
0
128
64
32
16
8
0
2
1
3