command:
:ROUTe[n][:CHANnel[m]]:CONFig?
syntax:
:ROUTe[n]:[CHANnel[m]]:CONFig?
description:
Queries the switch configuration of the instrument. For each channel, the minimum and maximum channel number of each
port is given.
parameters:
none
response:
<j>,<k>;<l>,<m> as text string where:
<j> is the first port character on the left
<k> is the last port character on the left
<l> is the minmimum port number on the right
<m> is the maximum port number on the right
example:
rout2:conf?
affects:
All switch modules
command:
:ROUTe[n][:CHANnel[m]]:CONFig:ROUTe?
syntax:
:ROUTe[n]:[CHANnel[m]]:CONFig:ROUTe?
description:
Queries the allowed switch routes of an instrument.
parameters:
none
response:
[A.....Z],[1.....n];[A.....Z],[1.....n].[A.....Z],[1.....n] as a text string.
"," separates input and output ports of a single connection.
";" separates parallel connections
"." separates possible switch states
example:
rout2:conf:rout?
2x2 x-over switcch:
state 1: When A to 1 then B to 2nd connection (straight)
state 2: When A to 2 then B to 1st connection (cross-over)
affects:
All switch modules
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
->
A,B;1,2 (2 left and 2 right ports)
->
A,1;B,2.A,2;B,1
Measurement Operations & Settings