Keysight Technologies 8163A/B Programming Manual page 114

Table of Contents

Advertisement

4
Measurement Operations & Settings
command:
:SENSe[n]:[CHANnel[m]]:POWer:REFerence:STATe:RATio
syntax:
:SENSe[n]:[CHANnel[m]]POWer:REFerence:STATe:RATio<wsp>
<slot number>|255|TOREF,<channel number>
description:
Selects the reference for the module.
parameters:
slot number:
255 or TOREF:
channel number:
If you want to reference another power sensor channel, use an integer value corresponding to the slot for the first parameter
and an integer value corresponding to the channel for the second value.
If you want to use an absolute reference, use TOREF as the first parameter and any integer value as the second parameter.
response:
none
examples:
sens1:pow:ref:stat:rat 2,1
sens1:pow:ref:stat:rat TOREF,1
affects:
All power meters
dual sensors:
Master and slave channels are independent.
command:
:SENSe[n]:[CHANnel[m]]:POWer:REFerence:STATe:RATio?
syntax:
:SENSe[n]:[CHANnel[m]]POWer:REFerence:STATe:RATio?
description:
Returns the reference setting for the module.
parameters:
none
response:
results are displayed relative to an absolute reference or to the current power reading from another channel.
examples:
sens1:pow:ref:stat:rat?
sens1:pow:ref:stat:rat?
affects:
All power meters
dual sensors:
Master and slave channels are independent.
114
an integer value representing the slot number you want to reference
results are displayed relative to an absolute reference
an integer value representing the channel number you want to reference
->
+255,+0<END>
->
+2,+1<END>
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
References channel 2.1
References an absolute reference
results are displayed relative to an
absolute reference
results are displayed relative to channel
2.1