Keysight Technologies 8163A/B Programming Manual page 138

Table of Contents

Advertisement

4
Measurement Operations & Settings
command:
[:SOURce[n]][:CHANnel[m]]:FREQuency:TOGRid
syntax:
[:SOURce[n]][:CHANnel[m]]:FREQuency:TOGRid <wsp>
description:
Set the frequency to the nearest grid point only changing the channel number.
parameters:
The frequency value.
response:
None
example:
sour4:freq:togr 188.5thz
affects:
81950A
Notes
Only available in "Grid Mode"
The frequency offset is neither changed nor taken into account when calculating the nearest channel.
command:
[:SOURce[n]][:CHANnel[m]]:FREQuency:AUTO
syntax:
[:SOURce[n]][:CHANnel[m]]:FREQuency:AUTO<wsp>OFF|0|ON|1
description:
Switch between "Auto Mode" and "Grid Mode"
parameters:
OFF or 0: "Grid Mode" (= Auto Mode off)
ON or 1: "Auto Mode"
response:
None
example:
sour1:freq:auto on
affects:
81950A
Notes
This command is equivalent to [:SOURce[n][:CHANnel[m]]:WAVelength:AUTO
138
<value>[HZ|KHZ|MHZ|GHZ|THZ]