Keysight Technologies 8163A/B Programming Manual page 177

Table of Contents

Advertisement

parameters:
0 or STOP:
1 or STARt:
2 or PAUSe:
3 or CONTinue:
If you enable lambda logging (see
page 170 ) and modulation (see
sweep cannot be started.
Generally, a continuous sweep can only be started if:
the trigger frequency, derived from the sweep speed and sweep step, is <= 40kHz, or <=1MHz for 81602A, 81606A, 81607A,
81608A, and 81960A.
the number of triggers, calculated from the sweep span and sweep span, is <=100001
the start wavelength is less than the stop wavelength.
In addition, a continuous sweep with lambda logging requires:
the trigger output to be set to step finished
modulation set to coherence control or off.
response:
none
example:
wav:swe STOP
affects:
All tunable laser modules, except 81950A.
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
Stop the sweep.
Start a sweep, run sweep.
Pause the sweep. (doesn't apply for continuous sweep)
Continue a sweep. (doesn't apply for continuous sweep)
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:LLOGging
[:SOURce[n]][:CHANnel[m]]:AM:STATe[l]
Measurement Operations & Settings
on page 127 ) simultaneously, a
4
on
177