Keysight Technologies 8163A/B Programming Manual page 162

Table of Contents

Advertisement

4
Measurement Operations & Settings
command:
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ZERO:TEMPerature:DIFFerence?
syntax:
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ZERO:TEMPerature:Difference?
description:
Reports the temperature difference required to trigger an auto lamda zero.
parameters:
none
response:
float value; temperature in °C
example:
sour0:wav:corr:zero:temp:diff?
affects:
All tunable laser modules except 81649A, 81689A/B and 819xxA/B, 81602A, 81606A, 81607A, 81608A, 81609A, and
81950A.
command:
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ZERO:TEMPerature:LASTzero?
syntax:
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ZERO:TEMPerature:LASTzero?
description:
Reports the temperature at which the last auto lamda zero took place.
parameters:
none
response:
float value; temperature in °C
example:
sour0:wav:corr:zero:temp:last?
affects:
All tunable laser modules except 81649A, 81689A/B and 819xxA/B, 81602A, 81606A, 81607A, 81608A, 81609A, and
81950A.
command:
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ZERO:AUTO
syntax:
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ZERO:AUTO
description:
Forces an auto lamda zero. This is quicker but a little less accurate than the equilavent manual process because some checks
are omitted:
parameters:
none
response:
none
example:
sour0:wav:corr:zero:auto
affects:
All tunable laser modules except 81649A, 81689A/B and 819xxA/B, 81602A, 81606A, 81607A, 81608A, 81609A. and
81950A.
162
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

8164a/b8166a/b