-200 to -299 Execution Errors
Standard
New
New
Old
Old
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
Error
Number
String
-185
"Subop out of range"
Description:
Suboperations are parameters that are passed to refine the destination of a command.
They are used to address slots, channels, laser selections and GPIB/SCPI register levels.
This error is generated if the parameter is not valid in the current context or system
configuration.
Example:
This error occurs if the user queries the status of a summary register and passes an
invalid status level (also see "Status for 816x" on page 28 programmer's guide).
Note:
Incorrect slots and channels addresses are handled by error code -301
-200
"Execution error (StatExecError)"
Description:
This error occurs when the current function, instrument or module state (or status)
prevents the execution of a command. This is a generic error which can occur for a
number of reasons.
Example:
When a powermeter has finished a logging application and data is available, the user is
not able to reconfigure the logging application parameters. First, the user must stop the
logging application.
-201
"Please be patient - GPIB currently locked out"
Description:
Some operations block the complete system. Since no sensible measurements are
possible while this is true, the GPIB is locked out.
Example:
When ARA, Lambda zeroing or zeroing is executing on a TLS module, the GPIB is not
accessible.
-205
"Powermeter not running (StatMeterNotRunning)"
Description:
Some command and actions may stop the data aquisition unit of a powermeter. If a
command fetches data, there may be no measurement values and this error is generated.
Please check module state and repeat operation.
-211
"Trigger ignored"
Description:
A trigger has been detected but ignored because of timing contraints. (For Example:
average time to large).
-212
"Arm ignored"
Description:
The user can set the automatic re-arming option for input and output trigger events (see