Noise figure meter security and memory clearing

Advertisement

Quick Links

Keysight Technologies
8970B Noise Figure Meter
Security and
Memory Clearing

Advertisement

loading

Summary of Contents for Keysight Technologies 8970B

  • Page 1 Keysight Technologies 8970B Noise Figure Meter Security and Memory Clearing...
  • Page 2 The information contained in this document is subject to change without notice. Keysight Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 3: Tools Required

    8970B Noise Figure Meter Security and Memory Clearing 8970B Noise Figure Meter Security and Memory Clearing Introduction 8970B memory clearing process requires removal of the processor battery for 15 minutes or longer. Below are step-by-step instructions for removing this battery. Make sure you use proper ESD pre-cautions.
  • Page 4: Contacting Keysight

    13.You may now calibrate the 8970B as desired for your next measurement. Contacting Keysight For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your local Keysight office access the following URL, or if in the United States, call the following telephone number: http://www.keysight.com/find/assist...