8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
Especially when using VXI-11 connections, lock the device.
Otherwise one controller may get the reply for another controller's
query.
If the device is locked in a VISA socket connection, you can still
access it using raw sockets (that is, through directly using sockets).
When using sockets (VISA and raw) there are some extra points to be
aware of:
Always append a 'newline' to your command or query.
Don't use indefinite blocks.
It is the programmer's responsibility to make sure all the transmitted
data is read.
The Agilent VISA Assistant is confused by 'newline' inside a reply. In
this case, click the button [viScanf] until the whole reply is read.
VXI-11 connections are about 50% slower than Socket