Keysight Technologies 8163A/B Programming Manual page 56

Table of Contents

Advertisement

3
Instrument Setup and Status
command:
:STATus:OPERation:CONDition[:LEVel0]?
syntax:
:STATus:OPERation:CONDition[:LEVel0]?
description:
Reads the Operational Status Condition Summary Register.
parameters:
none
response:
The sum of the results for the individual slots (a 16-bit signed integer value, where 0
Bits
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
example:
stat:oper:cond?
56
Mnemonics
8163A/B
8164A/B
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Slot 4 Summary
Not used
Slot 3 Summary
Slot 2 Summary
Slot 2 Summary
Slot 1 Summary
Slot 1 Summary
Not used
Slot 0 Summary
->
+0<END>
8166A/B
Not used
Slot 14 Summary
Slot 13 Summary
Slot 12 Summary
Slot 11 Summary
Slot 10 Summary
Slot 9 Summary
Slot 8 Summary
Slot 7 Summary
Slot 6 Summary
Slot 5 Summary
Slot 4 Summary
Slot 3 Summary
Slot 2 Summary
Slot 1 Summary
Level 1 Summary
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
value
32767):
Decimal Value
0
16384
8192
4096
2048
1024
512
256
128
64
32
16
8
4
2
1

Advertisement

Table of Contents
loading

This manual is also suitable for:

8164a/b8166a/b