Keysight Technologies 8163A/B Programming Manual page 53

Table of Contents

Advertisement

command:
*TST?
syntax:
*TST?
description:
The self-TeST query *TST? makes the instrument perform a self-test and place the results of the test in the output queue. If
the self-test fails, the results are also put in the error queue. We recommend that you read self-test results from the error
queue. No further commands are allowed while the test is running. After the self-test the instrument is returned to the
setting that was active at the time the self-test query was processed. The self-test does not require operator interaction
beyond sending the *TST? query.
parameters:
none
response:
The sum of the results for the individual tests (a 32-bit signed integer value, where 0
Bits
31
18 - 30
17
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
Mnemonic
Selftest failed on Mainframe
Not used
Selftest failed on Slot 17
Selftest failed on Slot 16
Selftest failed on Slot 15
Selftest failed on Slot 14
Selftest failed on Slot 13
Selftest failed on Slot 12
Selftest failed on Slot 11
Selftest failed on Slot 10
Selftest failed on Slot 9
Selftest failed on Slot 8
Selftest failed on Slot 7
Selftest failed on Slot 6
Selftest failed on Slot 5
Selftest failed on Slot 4
Selftest failed on Slot 3
Selftest failed on Slot 2
Selftest failed on Slot 1
Instrument Setup and Status
value
4294967296):
Decimal Value
A negative value
0
131072
65536
32768
16384
8192
4096
2048
1024
512
256
128
64
32
16
8
4
2
3
53

Advertisement

Table of Contents
loading

This manual is also suitable for:

8164a/b8166a/b

Table of Contents