Hard ware
Trigger Rearming
Triggering
trig:outp
trig:outp:rearm
DISabled
-
AVGover
ON
MEASure
ON
AVGover
OFF
MEASure
OFF
command:
:TRIGger
syntax:
:TRIGger<wsp>NODEA|1|NODEB|2
description:
Generates a hardware trigger.
parameters:
1 or NODEA:
2 or NODEB:
A hard ware trigger cannot be effective in the DISabled triggering mode but can be effective in DEFault, PASSthrough or
LOOPback triggering modes, see
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
Table 9
Generating Output Triggers from Power Measurements
Software Triggering
init:imm
init:cont
An output trigger will never be generated.
An output trigger is generated for every new power measurement when the averaging time period
finishes.
An output trigger is generated for every new power measurement when the averaging time period begins.
An output trigger is generated when the averaging time period of
the first power measurement finishes. A further hard ware output
trigger cannot be generated until you send trig:outp:rearm.
An output trigger is generated when the averaging time period of
the first power measurement begins. A further hard ware output
trigger cannot be generated until you send trig:outp:rearm.
Is identical to a trigger at the Input Trigger Connector.
Generates trigger at the Output Trigger Connector.
:TRIGger:CONF
Measurement Operations & Settings
Data Acquisition Functions
sens:func:stat
MINMax
on page 214 for information on triggering modes.
LOGGing|STABility
Applies for all subsequent data
acquisition functions.
Applies for all subsequent data
acquisition functions.
An output trigger is generated for
every new power measurement
when the averaging time period
finishes. Applies for all subsequent
data acquisition functions.
An output trigger is generated for
every new power measurement
when the averaging time period
begins. Applies for all subsequent
data acquisition functions.