Keysight Technologies 8163A/B Programming Manual page 118

Table of Contents

Advertisement

4
Measurement Operations & Settings
response:
none
example
sens1:ret:cal:coll:refl
affects:
All return loss modules
command:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:TERMination
syntax:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:TERMination
description:
For the currently selected source, start the calibration and save the calibration values for a defined termination reference
measurement. See
information on setting the return loss value of your reference reflector.
parameters:
none
response:
none
example
sens1:ret:cal:coll:term
affects:
All return loss modules
command:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:TERMination?
syntax:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:TERMination?
description:
Queries the T-value (termination calibration value) for the return loss module
parameters:
none
response:
Termination calibration value as a float in dB
example
sens1:ret:cal:term?
affects:
All return loss modules
118
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l]
->
+6.5000E+001
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
on page 121 for

Advertisement

Table of Contents
loading

This manual is also suitable for:

8164a/b8166a/b

Table of Contents