Keysight Technologies 85059B Operating And Service Manual
Keysight Technologies 85059B Operating And Service Manual

Keysight Technologies 85059B Operating And Service Manual

1.0 mm precision calibration kit
Table of Contents

Advertisement

Quick Links

Keysight Technologies
85059B
1.0 mm Precision
Calibration Kit
Operating and
Service Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 85059B and is the answer not in the manual?

Questions and answers

Summary of Contents for Keysight Technologies 85059B

  • Page 1 Operating and Service Manual Keysight Technologies 85059B 1.0 mm Precision Calibration Kit...
  • Page 2 DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY beyond those set forth in the TERMS IN THE SEPARATE EULA shall apply, except to the © Keysight Technologies, Inc. AGREEMENT WILL CONTROL. extent that those terms, rights, or 2016-2017 licenses are explicitly required...
  • Page 3: Table Of Contents

    85059B (Precision) Kit Contents ........
  • Page 4 Contents Mechanical Characteristics ........... 2-3 Electrical Specifications.
  • Page 5 Contents Where to Look for More Information ......... . . 5-3 Returning a Kit or Device to Keysight .
  • Page 6 Contents Keysight 85059V Operation and Service Guide...
  • Page 7: General Information

    85059B Operation and Service Guide General Information Chapter One at-a-Glance About this Manual This manual describes the 85059B calibration kit and provides replacement part numbers, specifications, and procedures for using, maintaining and troubleshooting this kit. Chapter One at-a-Glance Section Title Summary of Content “Calibration Kit Overview”...
  • Page 8: Section Title

    General Information Chapter One at-a-Glance Section Title Summary of Content “Regulatory and Environmental Description of regulatory symbols. Information” on page 1-15 Keysight 85059B Operation and Service Guide...
  • Page 9: Calibration Kit Overview

    General Information Calibration Kit Overview Calibration Kit Overview The 85059B 1.0 mm precision calibration kit is used to calibrate your PNA series network analyzer system. This kit is used for the measurement of Ω components with 50 , 1.0 mm connectors with a frequency range of DC to 120 GHz.
  • Page 10: Data-Based Model

    In the PNA’s Modify Calibration Class Assignments dialog box , the “Expanded Math” feature is turned ON by default when the 85059B data-based model is selected from a cal kit menu. In order to achieve a more accurate calibration, Expanded Math uses the data at any given frequency from ALL measured standards instead of using only the three standards specified in the class definitions.
  • Page 11: Offset Opens And Shorts

    1-1). Figure 1-1 OPEN Calibration Devices: 85059B Adapters Like the other devices in the kit, the adapters are built to very tight tolerances to provide good broadband performance and to ensure stable, repeatable connections. Keysight 85059B Operation and Service Guide...
  • Page 12: Lowband Loads

    Equipment Required but Not Supplied Various connector cleaning supplies and electrostatic discharge safety supplies are not provided in this kit. (Refer to Chapter 6, “Replacement Parts”, on page 6- 1 for ordering information.) Keysight 85059B Operation and Service Guide...
  • Page 13: Power Calibration

    Refer to “Using a Network Analyzer” on page 1-6. To download a copy of the 85059B calibration kit definitions (.xkt file), refer to Keysight’s Calibration Kit Definitions Web page at http://na.support.keysight.com/pna/caldefs/stddefs.html For best measurement results, the isolation standard should be the equivalent impedance of the DUT.
  • Page 14: Calibration Residual Error Specifications

    General Information Calibration Residual Error Specifications Calibration Residual Error Specifications Please download our free Vector Network Analyzer Uncertainty Calculator from http://www.keysight.com/find/na_calculator. Table 1-1 85059B (Precision) Residual System Calibration Error Specifications Frequency Range (GHz) Data-Based Model (Residuals) Directivity Source Reflection Match...
  • Page 15: Serial Numbers

    (serial numbers are printed on an attached label, or scribed onto the body of each device). Record these serial numbers in Figure 1-6 on page 1-16. This can help you avoid confusing the devices in this kit with similar devices from other kits. Keysight 85059B Operation and Service Guide...
  • Page 16: Recording The Device Serial Numbers

    General Information Incoming Inspection Recording the Device Serial Numbers Table 1-2 Serial Number Record for 85059B (Precision) Device Serial Number Calibration kit _______________________________ Load –m– _______________________________ Load –f– _______________________________ Offset open –m– _______________________________ Offset open –f– _______________________________ 1.3 mm offset short 1 –m–...
  • Page 17 General Information Incoming Inspection — the part number and serial number of the device — the type of service required — a detailed description of the problem Keysight 85059B Operation and Service Guide 1-11...
  • Page 18: Clarifying Connector Gender

    In this manual, connectors are referred to in terms of their device gender unless otherwise stated. For example, a male open has a male connector. Figure 1-3 Clarifying Connector Gender Figure 1-4 Male and Female Connectors 1-12 Keysight 85059B Operation and Service Guide...
  • Page 19: Clarifying The Terminology Of A Connector Interface

    Improper connections, resulting from pin depth values being out of the typical limits (see Table 2-4), or from bad connections, can also damage these devices. Keysight 85059B Operation and Service Guide 1-13...
  • Page 20: When To Calibrate

    With experience you will be able to see changes in the reference responses that indicate a need to perform the measurement calibration again. 1-14 Keysight 85059B Operation and Service Guide...
  • Page 21: Regulatory And Environmental Information

    The Keysight email address is required by EU directives applicable to our product. sight.com China RoHS regulations include requirements related to packaging, and require compliance to China standard GB18455-2001. This symbol indicates compliance with China RoHS regulations for paper/fiberboard packaging. Keysight 85059B Operation and Service Guide 1-15...
  • Page 22 General Information Regulatory and Environmental Information 1-16 Keysight 85059B Operation and Service Guide...
  • Page 23: Specifications

    Keysight 1.0 mm Precision Calibration Kit 85059B Operation and Service Guide Specifications Information in This Chapter Chapter Two at-a-Glance Section Title Summary of Content “Environmental Requirements” on Describes the operating and storage environmental conditions for your page 2 calibration kit.
  • Page 24: Environmental Requirements

    +19 °C to +21 °C. It is then appropriate to perform measurements and performance verifications at +19 °C, which is outside the operating temperature range of the calibration devices, since only the actual calibration must be performed within the operating temperature range. Keysight 85059B Operation and Service Guide...
  • Page 25: Mechanical Characteristics

    Technologies verifies the mechanical characteristics of the devices in this kit with special gaging processes and electrical testing. This ensures that the device connectors do not exhibit any improper pin depth when the kit leaves the factory. Keysight 85059B Operation and Service Guide...
  • Page 26: Electrical Specifications

    Electrical Specifications Electrical Specifications The electrical specifications in Table 2-2 apply to the devices in your calibration kit when connected with a Keysight precision interface. Table 2-2 85059B (Precision) Electrical Specifications for 1.0 mm Devices Device Frequency Parameter (GHz) Data-Based Model...
  • Page 27 Data-Based Model Device Frequency Parameter (GHz) Return Loss (dB) Insertion Loss (dB) (typical) Adapters DC to 20 —0.5 20 to 50 —0.5 50 to 75 —0.5 75 to 110 —0.5 110 to 120 —0.7 Keysight 85059B Operation and Service Guide...
  • Page 28: Certification

    Specifications Electrical Specifications Certification Keysight Technologies certifies that this product met its published specifications at the time of shipment from the factory. Keysight further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST) to the extent allowed by the institute’s calibration facility, and to the calibration facilities of other...
  • Page 29: Use, Maintenance, And Care Of The Devices

    Keysight 1.0 mm Precision Calibration Kit 85059B Operation and Service Guide Use, Maintenance, and Care of the Devices Information in this Chapter Chapter Three at-a-Glance Section Title Summary of Content “Electrostatic Discharge” on page 2 How to protect your instruments and devices against electrostatic discharge.
  • Page 30: Electrostatic Discharge

    3. Connect the other end of the cable to the test port. 4. Remove the short from the cable. Refer to Chapter 6, “Replacement Parts.” for part numbers and instructions for ordering ESD protection devices. Keysight 85059B Operation and Service Guide...
  • Page 31 Use, Maintenance, and Care of the Devices Electrostatic Discharge Figure 3-1 ESD Protection Setup Keysight 85059B Operation and Service Guide...
  • Page 32: Visual Inspection

    Connector nuts should move smoothly and be free of: — Burrs — Loose metal particles — Rough spots Any connector that has obvious defects should be discarded or sent for repair - refer to “Contacting Keysight” on page 5-4. Keysight 85059B Operation and Service Guide...
  • Page 33: What Causes Connector Wear

    Notice the location of the cross hairs in relationship to the center of the figures. Figure 3-2 Contact Integrity Concentricity Figure 3-3 Figure 3-4 show the concentricity of both the male and female 1.0 mm connectors. Inspect the connectors with a minimum magnification of 10X. Keysight 85059B Operation and Service Guide...
  • Page 34: Inspect The Mating Plane Surfaces

    None of these affect electrical or mechanical performance. Clean and inspect the connector again, if it shows: — Deep scratches or dents — Particles clinging to the mating plane surfaces — Uneven wear Keysight 85059B Operation and Service Guide...
  • Page 35: Inspect Female Connectors

    Inspection is particularly important when mating nonprecision to precision devices. Supplies and Equipment Needed The supplies and equipment needed to perform the cleaning procedure, and their Keysight Technologies part numbers are listed in Table 7-1 on page 7-2 and page 7-3. Keysight 85059B Operation and Service Guide...
  • Page 36: Cleaning Connectors

    Use a lint-free swab or cleaning cloth moistened with isopropyl alcohol to remove any dirt or stubborn contaminants on a connector that cannot be removed with compressed air or nitrogen. Refer to Table 6-2 on page 6-4 for a part numbers for cleaning swabs. Keysight 85059B Operation and Service Guide...
  • Page 37 Always completely dry a connector before you reassemble or use it. 4. Reinspect Inspect the connector again to make sure that no particles or residue are present. Figure 3-5 Cleaning Illustration Keysight 85059B Operation and Service Guide...
  • Page 38: Pin Depth

    Keysight verifies the pin depth characteristics of the connectors in this kit with special gaging processes and electrical testing. This ensures that the device connectors do not exhibit any center conductor protrusion and have proper pin depth when the kit leaves the factory. 3-10 Keysight 85059B Operation and Service Guide...
  • Page 39 −0.003 to −0.011 mm −0.00011 to −0.00043 in Shorts −0.003 to −0.007 mm −0.00011 to −0.00027 in Fixed Loads −0.0020 to −0.020 mm −0.000079 to −0.00078 in Adapters −0.004 to −0.024 mm −0.00016 to −0.00094 in Keysight 85059B Operation and Service Guide 3-11...
  • Page 40: Making Connections

    Use very light finger pressure (no more than 2 inch–pounds of torque). 7. Relieve any side pressure on the connection from long or heavy devices, or cables. This assures consistent torque (refer to “Using a Torque Wrench” on page 3-15). 3-12 Keysight 85059B Operation and Service Guide...
  • Page 41: Connector Misalignment

    Turn the connector nut, not the device body. Major damage to the center conductor can occur if the device body is twisted. 1. Use an open-end wrench to prevent the device body from turning. 2. Use another open-end wrench to loosen the connector nut. Keysight 85059B Operation and Service Guide 3-13...
  • Page 42 Use, Maintenance, and Care of the Devices Making Connections 3. Complete the separation by hand, turning only the connector nut. 4. Pull the connectors straight apart without twisting, rocking, or bending either of the connectors. 3-14 Keysight 85059B Operation and Service Guide...
  • Page 43: Using A Torque Wrench

    3-16). Wrenches opposing each other (180 degrees apart) will cause a lifting action. This lifting action can misalign, and stress the connections of the devices involved. This is especially true when several devices are connected together. Keysight 85059B Operation and Service Guide 3-15...
  • Page 44 Wide separation of the wrenches produces a larger residual lateral force on the structure of connected devices. This can degrade connector repeatability. 3. Hold the torque wrench lightly, at the end of the handle only (beyond the groove). See Figure 3-11 on page 3-17. 3-16 Keysight 85059B Operation and Service Guide...
  • Page 45 Using the Torque Wrench 5. Tighten the connection just to the torque wrench "break" point. The wrench handle gives way at its internal pivot point. See Figure 3-12. Do not tighten the connection further. Keysight 85059B Operation and Service Guide 3-17...
  • Page 46 Do not twist the head of the wrench relative to the outer conductor mating plane. If you do, you apply more than the recommended torque. 3-18 Keysight 85059B Operation and Service Guide...
  • Page 47: Disconnection Procedure

    — Do not set connectors contact-end down on a hard surface. The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface. Keysight 85059B Operation and Service Guide 3-19...
  • Page 48 Use, Maintenance, and Care of the Devices Handling and Storage 3-20 Keysight 85059B Operation and Service Guide...
  • Page 49: Performance Verification

    Introduction The performance of your calibration kit can only be verified by returning the kit to Keysight Technologies for recertification. The equipment required to verify the specifications of the devices in the kit has been specially manufactured and is not commercially available.
  • Page 50: How Keysight Verifies The Devices In This Kit

    These two steps establish a traceable link to NIST for Keysight to the extent allowed by the institute’s calibration facility. The specifications data provided for the devices in this kit is traceable to NIST through Keysight Technologies. Recertification The following will be provided with a recertified kit: —...
  • Page 51: How Often To Recertify

    The recertification interval should begin on the date the kit is first used after the recertification date. Where to Send a Kit for Recertification Contact Keysight Technologies for information on where to send your kit for recertification. See “Contacting Keysight” on page 5-4.
  • Page 52 Performance Verification Where to Send a Kit for Recertification Keysight 85059B Operation and Service Guide...
  • Page 53: Troubleshooting

    Keysight 1.0 mm Precision Calibration Kit 85059B Operation and Service Guide Troubleshooting Information in This Chapter Chapter Five at-a-Glance Section Title Summary of Content “Troubleshooting Process” on How to troubleshoot a calibration problem. Includes a flow page 5-2 troubleshooting diagram.
  • Page 54: Troubleshooting Process

    Troubleshooting Troubleshooting Process Troubleshooting Process If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow the steps in Figure 5-1. Figure 5-1 Troubleshooting Flowchart Keysight 85059B Operation and Service Guide...
  • Page 55: Where To Look For More Information

    “Contacting Keysight” on page 5-4. Returning a Kit or Device to Keysight If your kit or device requires service, contact Keysight Technologies for information on where to send it and proved the following: — your company name and address — a technical contact person within your company, and the person's complete phone number —...
  • Page 56: Contacting Keysight

    Keysight web site: — Go to www.keysight.com. — Enter the document’s part number (located on the title page) in the Search box. — Click Search — Open the PDF and print the document. Keysight 85059B Operation and Service Guide...
  • Page 57: Replacement Parts

    6-4) Introduction The following replaceable parts table lists the replacement part numbers for the 85059B calibration kit contents. To order a listed part, note the description, part number, and the quantity desired and contact Keysight Technologies - refer to “Contacting Keysight” on page...
  • Page 58 Replacement Parts Introduction Table 6-1 Replaceable Parts for the 85059B (Precision) Calibration Kit Description Qty Per Kit Keysight Part Number Calibration Devices (1.0 mm) 1.3 mm offset short 1 –m– 85059-60027 2.45 mm offset short 2 –m– 85059-60029 3.326 mm offset short 3 –m–...
  • Page 59 “Printing Copies of Documentation from the Web” on page 5-4. Figure 6-1 Replaceable Parts for the 85059B (Precision) Calibration Kit (not all items are shown below. For a complete list, refer to See “Replaceable Parts for the 85059B (Precision) Calibration Kit” on page...
  • Page 60 Foam tipped cleaning swabs 9301-1243 a. See “Clarifying the Terminology of a Connector Interface” on page 1-13. b. Ruggedized connector. c. Keysight can no longer safely ship isopropyl alcohol, so customers should purchase it locally. Keysight 85059B Operation and Service Guide...
  • Page 61: Standard Definitions

    Keysight 1.0 mm Precision Calibration Kit 85059B Operation and Service Guide Standard Definitions Information in This Chapter Chapter Seven at-a-Glance Section Title Summary of Content Class Assignments and Standard Discussion on standard definitions and where to find more information on the Definitions Values are Available on web.
  • Page 62 Standard Definitions Class Assignments and Standard Definitions Values are Available on the Web Keysight 85059B Operation and Service Guide...
  • Page 63 85059B cal kit calibration certification, specifications entering characteristics permanently stored mechanical deviation from nominal phase, 85059B supplemental device class assignments conductor downloading from Keysight Web site mating plane 3-10 cleaning connectors connecting 3-12...
  • Page 64 3-13 inspection downloading class assignments & std definitions damage from the Web defects female connectors electrical specifications, 85059B incoming 1-10 electrostatic discharge mating plane supplies visual part numbers isopropyl alcohol, as cleaning solvent when making connections...
  • Page 65 Index open-end wrench 3-13 frequency, 85059B part number humidity opens pin depth 3-11 return loss, 85059B temperature parts specified frequency performance required but not supplied standard definitions performance verification downloading from Keysight Web site fail standards pin depth 3-10 NIST...
  • Page 66 Index part numbers 85059V Index...
  • Page 67 Installation Note Xxxxx-xxxxx...
  • Page 68 This information is subject to change without notice. © Keysight Technologies 2016-2017 Edition 1, August 2017 85059-90005 www.keysight.com...

Table of Contents