Keysight Technologies 8163A/B Programming Manual page 110

Table of Contents

Advertisement

4
Measurement Operations & Settings
example:
sens1:pow:rang:auto?
affects:
All power meters and return loss modules
dual sensors:
Can only be sent to master channel, slave channel parameters are identical.
command:
:SENSe[n]:[CHANnel[m]]:POWer:REFerence
syntax:
:SENSe[n]:[CHANnel[m]]:POWer:REFerence<wsp>
TOMODule|TOREF,<value>PW|NW|UW|MW|Watt|DBM|DB|MDB
description:
Sets the sensor reference value.
parameters:
TOMODule:
TOREF:
The reference as a float value.
You must append a unit type
• dB if you use TOMODule or
• Watts or dBm if you use TOREF.
The two reference values are completely independent. When you change the reference mode using the command
:SENSe[n]:[CHANnel[m]]:POWer:REFerence:STATe:RATio
reference value entered for the selected reference mode.
response:
none
example:
sens1:pow:ref tomod,-40DB
affects:
All power meters
dual sensors:
Master and slave channels are independent.
command:
:SENSe[n]:[CHANnel[m]]:POWer:REFerence?
syntax:
:SENSe[n]:[CHANnel[m]]:POWer:REFerence?<wsp>TOMODule|TOREF
description:
Returns the sensor reference value.
parameters:
TOMODule:
TOREF:
response:
The reference as a float value.
110
->
1<END>
Sets the reference value in dB used if you choose measurement relative to another channel
Sets the reference value in Watts or dBm if you choose measurement relative to a constant
reference value
Returns the reference value in dB used if you choose measurement relative to another channel
Returns the reference value in Watts or dBm if you choose measurement relative to a constant
reference value
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
on page 114, the instrument uses the last

Advertisement

Table of Contents
loading

This manual is also suitable for:

8164a/b8166a/b