response:
4 byte Intel floating point; the averaging time in seconds
example:
OUTP1:ATIM?
affects:
Attenuator modules with power control.
command:
:OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO
syntax:
:OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO
description:
Zeros the electrical offsets of the attenuator's integrated powermeter.
parameters:
none
response:
none
example:
OUTP1:CORR:COLL:ZERO
affects:
Attenuator modules with power control.
command:
:OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO:ALL
syntax:
:OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO:ALL
description:
Zero all available powermeter channels in the mainframe.
parameters:
none
response:
none
example:
OUTP1:CORR:COLL:ZERO:ALL
affects:
Powermeter modules; attenuator modules with power control, and return loss modules.
command:
:OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO?
syntax:
:OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO?
description:
Queries the status of the last :OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO operation.
parameters:
none
response:
integer
example:
OUTP:CORR:COLL:ZER0?
affects:
Attenuator modules with power control.
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
->
1<END>
0 = OK, otherwise not OK.
->
0<END>
Measurement Operations & Settings