command:
:CONFigure[n][:CHANnel[m]]:OFFSet:WAVelength:REFerence?
syntax:
:CONFigure[n]][:CHANnel[m]]:OFFSet:WAVelength:REFerence?
description:
Queries the currently selected slot and channel of the external powermeter (hosted in the same mainframe as the attenuator
module) used by TOREF.
See:
parameters:
none
response:
the slot and channel of the external powermeter as integer values.
example:
CONF1:OFFS:WAV:REF?
affects:
Attenuator modules with power control.
command:
:CONFigure[n][:CHANnel[m]]:OFFSet:WAVelength:VALue:WAVelength?
syntax:
:CONFigure[n]][:CHANnel[m]]:OFFSet:WAVelength:VALue:WAVelength?<wsp><index>
description:
Queries a wavelength value from its position, or index, in the offset table. Offset table entries are ordered from shortest to
longest wavelength. The first index number = 1.
parameters:
<index>
response:
4 byte Intel floating point; the wavelength in meters
example:
CONF1:OFFS:WAV:VAL:WAV? 1
affects:
All attenuator modules.
command:
:CONFigure[n][:CHANnel[m]]:OFFSet:WAVelength:VALue:OFFSet?
syntax:
:CONFigure[n]][:CHANnel[m]]:OFFSet:WAVelength:VALue:OFFSet?<wsp><index | wavelength [PM | NM | UM | MM| M],>
description:
Queries an offset value from the position, or index, of the associated wavelength in the offset table. Offset table entries are
ordered from shortest to longest wavelength. The first index number =1.
Or: Queries the offset applied for a particular wavelength.
parameters:
<index>
<wavelength>
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
:CONFigure[n][:CHANnel[m]]:OFFSet:WAVelength:VALue
->
+2,+1<END>
The position of the wavelength value in the offset table.
->
The position of the value pair (wavelength; offset) in the offset table.
The wavelength for the offset table entry, in m
+1.55000000E-006<END>
Measurement Operations & Settings
on page 198
4
199