Keysight Technologies 8163A/B Programming Manual page 294

Table of Contents

Advertisement

9
Error Codes
New/Old/Standard
Old
Old
Old
Old
294
Error
Number
String
-213
"Init ignored"
Description:
The INIT:IMM command
cycle. The continuous measurement must be DISABLED. This error code is generated if
the powermeter is still in cont. measurement mode.
-220
"Parameter error (StatParmError)"
Description:
The user has passed a parameter that cannot be changed in this way. The device cannot
detect one of the following more specific errors:
-220
-220, "Parameter error (StatParmOutOfRange)"
Description:
The user has passed a parameter that exceeds the valid range for this parameter.
-220
"Parameter error (StatParmIllegalVal)"
Description:
The user has passed a parameter that does not match a value in a list of possible values.
(page
89) initiates a trigger and completes a full measurement
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

8164a/b8166a/b

Table of Contents