command:
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:STEP:[WIDTh]
syntax:
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:STEP:[WIDTh]<wsp>
<value>[PM|NM|UM|MM|M]
description:
Sets the width of the sweep step.
In continuous sweep mode, the end of a step is used for triggering.
parameters:
The width of the sweep step as a float value.
If you specify no units in your command, meters are used as the default.
response:
none
example:
wav:swe:step 5nm
affects:
All tunable laser modules, except 81950A.
command:
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:STEP:[WIDTh]?
syntax:
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:STEP:[WIDTh]?[<wsp>MIN|MAX]
description:
Returns the width of the sweep step
parameters:
optional
response:
The sweep step as a float value in meters.
example:
wav:swe:step?
affects:
All tunable laser modules, except 81950A.
command:
:SOURce<SlotNr>:TRANsmitter:RECalibration
syntax:
:SOURce<SlotNr>:TRANsmitter:RECalibration
description:
Recalibrates a reference transmitter module.
The result can be queried using
It is either '+0,"No error"' or e.g. '-200,"Execution error (StatExecError)"'
Recalibration takes about 6.8 seconds, so it also takes this time to get the reply to the :SYSTem:ERRor? query.
parameters:
none
response:
none
example:
:SOUR1:TRAN:REC
affects:
81490A
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
MIN Returns the minimum step width available.
MAX Returns the maximum step width available.
->
+5.00000000E-009<END>