command:
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:CHECkparams?
syntax:
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:CHECkparams?
description:
Returns whether the currently set sweep parameters (sweep mode, sweep start, stop, width, etc.) are consistent. If there is a
sweep configuration problem, the laser source is not able to pass a wavelength sweep.
parameters:
none
response:
A string with a detailed description of a configuration problem, or "OK" if the sweep os configured correctly. The responses
shown below are all the possible configuration problem strings:
Message
368,LambdaStop <=LambdaStart
369,sweepTime < min
370,sweepTime > max
371,triggerFreq > max
372,step < min
373,triggerNum > max
374,LambdaLogging = On AND
Modulation = On AND
ModulationSource! =
CoherenceControl
375,LambdaLogging = On AND
TriggerOut! = StepFinished
376,Lambda logging in stepped
mode
377,step not multiple of <x>
378, triggerFreq < min
379, Continuous Sweep AND
Modulation = On
example:
sour0:wav:swe:chec?
affects:
All tunable laser modules except Keysight 81649A, Keysight 81689A/B, and 81950A.
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
Description
start wavelength must be smaller than stop wavelength
the total time of the sweep is too small
the total time of the sweep is too large
the trigger frequency (calculated from sweep speed divided by sweep step) is too
large
step size too small
the number of triggers exceeds the allowed limit
The only allowed modulation source with the lambda logging function is coherence
control.
lambda logging only works "Step Finished" output trigger configuration
lambda logging can only be done in continuous sweep mode
the step size must be a multiple of the smallest possible step size
the number of triggers exceeds the allowed limit
379,Continuous Sweep = On for 81602A, 81606A, 81607A, 81608A, 81609A
->
"triggerNum > max"