4
Measurement Operations & Settings
response:
none
example:
INP1:OFFS:DISP
affects:
All attenuator modules
command:
:INPut[n][:CHANnel[m]]:OFFSet:POWermeter
syntax:
:INPut[n][:CHANnel[m]]:OFFSet:POWermeter<wsp><slot>,<channel>
description:
Sets the offset factor (a
same mainframe) (P
a
parameters:
<slot>
<channel>
response:
none
example:
INP1:OFFS:POW 4,2
affects:
Attenuator modules with power control.
command:
:INPut[n][:CHANnel[m]]:ATTenuation:SPEed
syntax:
:INPut[n][:CHANnel[m]]:ATTenuation:SPEed<wsp><value> | MIN | MAX | DEF
description:
Sets the filter transition speed; the speed at which the module moves from one attenuation to another (in dBs).
parameters:
<value>
MIN | MAX | DEF
response:
none
example:
INP1:ATT:SPE 2
affects:
All attenuator modules.
command:
:INPut[n][:CHANnel[m]]:ATTenuation:SPEed?
syntax:
:INPut[n][:CHANnel[m]]:ATTenuation:SPEed?<wsp> MIN | MAX | DEF
description:
Without the optional parameter, queries the transition speed of the filter.
parameters:
MIN | MAX | DEF
184
) to the difference between a power value measured by another powermeter (hosted by the
Offset
) and the power value measured by the attenuator module's monitor diode (P
ext
(dB) = P
(dBm) - P
(dBm)
Offset
att
ext
Slot number of the external powermeter.
Channel number of the external powermeter.
The filter transition speed in dB/s.
Sets the filter transition speed to the module limits, or the module default.
Queries the transition speed limits, or the module default.