command:
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ARA:ALL
syntax:
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ARA:ALL
description:
Realigns the laser cavity of every tunable laser source in a mainframe.
parameters:
none
response:
none
example:
sour0:wav:corr:ara:all
affects:
All tunable laser modules except 81649A, 81689A/B, and 81950A
command:
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:AUTocalib
syntax:
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:AUTocalib<wsp>ON (1) | OFF (0)
description:
Sets the Auto Calibration feature On or OFF. Switching it OFF enables the TLS to operate for a long period without
interruption from the "auto lambda zeroing" or settling. When Auto Calibration is disabled, it is possible to operate the TLS at
a temperature that differs more than 4.4 K from the last Lambda Zeroing temperature. In this case, the accuracy and
wavelength performance of the TLS can become less optimal due to temperature variation. The relevent accuracy class is
indicated on the user interface when Auto Calibration is off.
parameters:
a boolean value: 1 or ON: enable Autocalibration
response:
none
example:
sour0:wav:corr:aut 0
affects:
All tunable laser modules except 81649A, 81689A/B and 81980A, 81940A, 81989A, 81949A, 81602A, 81606A, 81607A,
81608A, 81609A, and 81950A.
command:
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:AUTocalib?
syntax:
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:AUTocalib?
description:
Returns whether Autocalibration is enabled or disabled
parameters:
none
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
0 or OFF: disable Autocalibration