Page 1
Reference Manual 00809-0100-4145, Rev AB September 2019 ™ Rosemount Smart Pressure Gauge...
Page 2
Changes or modifications to the equipment not expressly approved by Rosemount Inc. could void the user’s authority to operate the equipment. Using the Rosemount Smart Pressure Gauge in a manner other than what is specified by the manufacturer may impair the protection provided by the equipment.
Page 3
WARNING Physical access Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’ equipment. This could be intentional or unintentional and needs to be protected against. Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access by unauthorized personnel to protect end users’...
Field Communicator menu trees provides full menu trees and abbreviated fast key sequences for commissioning tasks. Models covered This manual covers the Rosemount Smart Pressure Gauge. • Measures gage/absolute/compound/vacuum pressure up to 10,000 psi (689.5 bar) Product recycling/disposal Recycling of equipment and packaging should be taken into consideration and disposed of in accordance with local and national legislation/regulations.
Page 8
Introduction Reference Manual September 2019 00809-0100-4145 Emerson.com/Rosemount...
The information in this section covers installation considerations. A Quick Start Guide is shipped with every device to describe basic installation and startup procedures. ™ Dimensional drawings for the Rosemount Smart Pressure Gauge can be found in the Product Data Sheet.
Page 10
The device needs to be turned on in order for the Field Communicator to interface with the Rosemount Smart Pressure Gauge. The Field Communicator connection is located to the right of the ON/OFF switch. To communicate with the device, connect the Field Communicator to connections labeled “COMM”.
Page 11
2.3.4 Electrical Battery The Rosemount Smart Pressure Gauge is self-powered. The battery contains approximately five grams of lithium. Under normal conditions, the battery materials are self-contained and are not reactive as the battery is maintained inside the enclosure of the device.
Use wrench on flats, not on housing. Mounting orientation The low side pressure port (atmospheric reference) on the pressure gauge is located in the neck of the device behind the housing. The vent path is between the housing and sensor. Figure 2-3. Emerson.com/Rosemount...
Page 13
During the power sequence, the dial tests full range of motion and LED flashes amber. 3. Once the power sequence ends, verify the LED flashes green. Note The LED may display several colors; see Table 4-2 for device statuses. Emerson.com/Rosemount...
Mount drain/vent valve upward to allow gases to vent. Gas flow measurement • Place taps in the top or side of the line. • Mount the device beside or above the taps so to drain liquid into the process line. Emerson.com/Rosemount...
A low side pressure port (atmospheric reference). WARNING Do not apply torque directly to the sensor module. Rotation between the sensor module and the process connection can damage the electronics. To avoid damage, apply torque only to the hex-shaped process connection. Emerson.com/Rosemount...
Hardware installation Reference Manual September 2019 00809-0100-4145 Rosemount manifolds The Rosemount 306 Integral Manifold mounts directly to the device. The manifold is used with this device to provide block-and-bleed valve capabilities of up to 10,000 psi (689.5 bar). 2.7.1 Installation procedure The Rosemount 306 Integral Manifold is for use only with a Rosemount Smart Pressure Gauge.
Page 17
2. To bring the device to atmospheric pressure, open the vent valve or bleed screw. Note A ¼-in. male NPT pipe plug may be installed in the test/vent port and will need to be removed with a wrench in order to vent the manifold properly. Emerson.com/Rosemount...
Page 18
3. After venting to atmosphere, perform any required calibration and then close the test/vent valve or replace the bleed screw. Device Test/vent (closed) Isolate Process (closed) 4. Open the Isolate (block) valve to return the device to service. Device Test/vent (closed) Isolate Process (open) Emerson.com/Rosemount...
Page 19
September 2019 Adjusting valve packing Over time, the packing material inside a Rosemount manifold may require adjustment in order to continue to provide proper pressure retention. Not all Rosemount manifolds have this adjustment capability. The Rosemount manifold model number will indicate what type of stem seal or packing material has been used.
Confirm correct device driver ™ Verify the latest Device Description (DD/DTM ) is loaded on your systems to ensure proper communications. Procedure 1. Visit the Emerson Device Install Kits Library or Fieldcommgroup.org. 2. Select desired product. a) Within Table 3-1, use the HART Universal Revision and Device Revision numbers to find the correct Device Description.
4. Follow the commands to perform the procedure. 3.5.2 Activate wireless Do not activate wireless until Emerson Wireless Gateway is installed and functioning properly; toggling off and on reduces battery life. Join device to network. 1. Obtain Network ID and Join Key for the wireless network (available in wireless gateway).
Reference Manual September 2019 00809-0100-4145 Note Devices are shipped from Emerson fully calibrated to the factory default of full scale (scale range = upper range limit). From the HOME screen, enter the Fast Key sequence. Fast Keys 2, 2, 1, 2 Procedure 1.
Page 25
From the HOME screen, enter the Fast Key sequence. Fast Keys 3, 2, 1 Procedure 1. From the Home screen, select 3: Service Tools. 2. Select 2: Variables. 3. Select 1: All Variables. The Operating Parameters menu displays the following information pertaining to the device: All variables: Emerson.com/Rosemount...
Wireless update rate From the HOME screen, enter the Fast Key sequence. Fast Keys 2, 2, 3, 2 1. From the Home screen, select 2: Configure. 2. Select 2: Advanced Setup. 3. Select 3: Wireless. 4. Select 2: Update Rate. Emerson.com/Rosemount...
Device reset The master reset function will reset the device electronics. To perform a device reset: From the HOME screen, enter the Fast Key sequence. Fast Keys 3, 3, 1 Procedure 1. From the Home screen, select 3: Service Tools. Emerson.com/Rosemount...
Page 29
<105% of MWP LED color: Green LED color: Green Dial location: Off-scale Dial location: Off-scale >105% MWP LED color: Green LED color: Red Dial location: Off-scale Dial location: Red X Local device status and notifications for more information. Emerson.com/Rosemount...
Page 30
From the HOME screen, enter the Fast Key sequence Fast Keys 3, 3, 3 Procedure 1. From the Home screen, select 3: Service Tools. 2. Select 3: Maintenance. 3. Select 3: Acknowledge Over-Pressure. 4. Follow the commands to perform the procedure. Emerson.com/Rosemount...
00809-0100-4145 September 2019 Operation and maintenance Overview ™ This section contains information on commissioning and operating Rosemount Wireless Pressure Gauges. Field Communicator and AMS Device Manager instructions are provided for convenience. Safety messages Procedures and instructions in this section may require special precautions to ensure the safety of the personnel performing the operation.
Page 32
Since this correction maintains the slope of the characterization curve, it should not be used in place of a sensor trim over the full sensor range. Emerson.com/Rosemount...
Page 33
Note Do not perform a zero trim on the Rosemount Smart Pressure Gauge with absolute measurement type. Zero trim uses a zero reference against ambient air pressure for gage, vacuum, and compound pressure devices, while absolute pressure devices reference absolute zero.
Page 34
13 percent of span. Dial adjustment only impacts needle position and does not impact sensor. Note Dial adjustment adjusts the position of the factory dial calibration. It is possible to degrade the performance of the gauge if the operation is done improperly or inaccurately. Emerson.com/Rosemount...
Page 35
From the HOME screen, enter the Fast Key sequence Fast Keys 3, 3, 2 Procedure 1. Select 3: Service Tools. 2. Select 3: Maintenance. 3. Select 2: Restore to Default Settings. 4. Follow the screen prompts to recall sensor and dial trim. Emerson.com/Rosemount...
WARNING The Rosemount Pressure Gauge shall be used only with the battery (00G45-9000-0001) supplied by Rosemount. This battery has been officially tested with the device as required by the I.S. standards during the assessment of the Rosemount Pressure Gauge. The battery is not replaceable in a hazardous location.
Page 37
Table 4-2: Status Descriptions (continued) LED color Device status Battery replacement required OR Device is malfunctioning No color No power, verify ON/OFF switch is in “on” position If the dial is pointing towards the red “X”, refer to Troubleshooting for more information. Emerson.com/Rosemount...
Page 38
Operation and maintenance Reference Manual September 2019 00809-0100-4145 Emerson.com/Rosemount...
Reference Manual Troubleshooting 00809-0100-4145 September 2019 Troubleshooting Service support To expedite the return process outside of the United States, contact the nearest Emerson representative. ™ Contact information for a regional Rosemount office is provided on the last page of this document.
Page 40
Troubleshooting Reference Manual September 2019 00809-0100-4145 Table 5-1: Interpreting Local Notifications (continued) LED color Dial location Device status Recommended action(s) Black, no color No power Verify ON/OFF switch is in “ON” position. Emerson.com/Rosemount...
2. Scroll as needed to the green menu bar and click Documents & Drawings. 3. Click Manuals & Guides. 4. Select the appropriate Quick Start Guide. Ordering information, specifications, and drawings To view current Rosemount Smart Pressure Gauge ordering information, specifications, and drawings, follow these steps: Procedure 1. Go to Emerson.com/Rosemount/Rosemount-Smart- Pressure-Gauge.
Page 42
Reference data Reference Manual September 2019 00809-0100-4145 Emerson.com/Rosemount...
5. Percent Range Quality 5. Maintenance 6. Custom Scale 7. Custom Scale Quality Zero 9. Device Informa on Device Informa on 1. Iden fica on 2. Revisions 3. Materials of Construc on 4. Security 5. Dial/Faceplate 6. Capabili es Emerson.com/Rosemount...
Page 44
6. Percent of Rng Quality Se ngs 7. Sensor Temp 8. Sensor Temp Quality 3. Acknowledge Overpressure 9. Sensor Temp Unit 4. Install New Ba ery 10. Supply Voltage 11. Supply Voltage Quality Simulate 1. Pressure 2. Sensor Temp 3. Supply Voltage Emerson.com/Rosemount...
Page 45
6. Lower Limit 7. Isolator Material 8. Fill Fluid Manifold Informa on 1. Process Connector 2. Connector Material 3. O-Ring Material 4. Drain/Vent Material Material Remote Seals 1. Number of Seals 2. Type 3. Isolator Material 4. Fill Fluid Emerson.com/Rosemount...
Page 47
Reference Manual 00809-0100-4145 September 2019 Emerson.com/Rosemount...
Page 48
Emerson Terms and Conditions of Sale are available upon request. The Emerson logo is a Facebook.com/Rosemount trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Youtube.com/user/RosemountMeasurement Emerson family of companies. All other marks are the property of their respective owners.