Run the System

Preconditions Before You Start

If the collision warning does not disappear after initialization, contact your Beckman
Coulter representative.

Camera Inspection Window

Important
The inspection window has to be kept minimized during operation. To prevent
interface errors, do not close or exit this window.
The camera inspection window always runs in the background. This window shows tube
images as they are captured by the camera in the tube inspection unit. The name of the cap
color and tube type identified are displayed at the bottom of this window.
Preconditions Before You Start
Caution
Always run the AutoMate system with all covers locked and the doors and interlock
doors closed and secured to prevent injury.
Before you start the application confirm that:
• The system has been completely cleared. All tubes have been removed from output
workplaces, from the conveyor belts, and from grippers of any robot or tube lift.
• The
Workplace numbers are correctly assigned to the corresponding workplace.
The waste container and the cap waste container are
• The label printer is on. Label rolls are loaded.
Enough secondary tubes and pipette tips are
• The recapper has enough Parafilm.
2-8
Sorting Drive software is running on the Sorting-Drive Server
empty.
available.
PC.