Figure 2.1 The PC is located behind this door
4
After the operating system has loaded, locate the AutoMate icon on the desktop. Use the
touch screen and tap the icon twice to start the main application.
5
The start screen appears prompting you to initialize. Select OK to confirm.
Note
6
Initialization begins. The progress and initialization status of each component is
displayed in a separate window. After successful initialization, the window closes
automatically.
B67471B
When the system is initializing, conveyor belt 01 completes a full rotation and all
tube positions on it are checked for available tubes. Any tube detected on conveyor
belt 01 is displayed as a yellow dot (meaning tube without bar code label) on
conveyor belt 01 on the screen. Tubes that are unexpected at a position are marked
yellow and automatically placed on the error rack by robot 2. The system
automatically scans for tubes on conveyor belt 1. This default setting can be
deactivated by a Beckman Coulter representative. If a tube was physically available
before initialization and is expected by the software but cannot be found on
conveyor belt 01, a message prompts you to locate the tube and remove it.
2
Run the System
Start the System
2-11