Beckman Coulter AutoMate 2500 Series Instructions For Use Manual page 115

Sample processing system
Hide thumbs Also See for AutoMate 2500 Series:
Table of Contents

Advertisement

5
Completely lift the cover sheet. Remove the tube holder from the removal position.
6
Move the tube holders along the conveyor belts manually. Confirm that all tube holders
move without scratching the white frame material. If a tube is in contact with the frame
material, confirm that the tube holder is set in its correct position.
7
At the turning points, confirm that the tube holders do not touch the top metal cover.
8
Turn on the AutoMate using the switch under the monitor.
Remove All Tube Holders
1
Turn off the AutoMate using the switch under the monitor.
2
Search for the reference tube holder with the black mark. Each conveyor belt has one
reference tube holder.
B67471B
User Maintenance
Conveyor Belt Maintenance
3
3-17