Laser Setting Window; Target Power Ranges In The Laser Setting Screen - Beckman Coulter CytoFLEX SRT Instructions For Use Manual

Cell sorter
Table of Contents

Advertisement

Figure 5.8 Laser Setting Window
1. Enable/Disable: Enables or disables the laser.
IMPORTANT
2. Target Power (mW): Used to change the laser target power.
NOTE
column in the QC report area) for range limits of each laser.
3.
IMPORTANT
4.
laser power in the Actual Power (mW) column of the Laser Setting window.
Table 5.1 Target Power Ranges in the Laser Setting Screen
Laser
405
488
561
638
C37808AC
The laser target power setting is not available to users.
The power detector has ±1 mW tolerance. Refer to
: Sets the laser target power setting.
The Actual Power readings are +/- 1 mW.
: Reads the current laser power before the flow cell assembly and displays the current
Power Range (mW)
90
50
30
100
Table 5.1
or
Max Power (mW)
120
60
40
120
Sorting
Configuring Acquisition Settings
Figure 2.2
(the target Power
5-41
5