Figure 5.6 Top view of the tube hopper with the slider assembly in its default position
4
Pull out the slider completely. Tubes start falling out.
5
Empty the tubes.
6
If the tube jam is cleared, skip step 7 to step 15. Continue with
7
If tubes are still jammed, depressurize the compressor.
Warning
B67471B
When the compressor is under pressure, the piston exerts force on the tube. To
avoid injury, do not try to remove any jammed tubes before depressurizing the
compressor.
Troubleshooting
Aliquot Module Troubleshooting
step
16.
5
5-15