Replace the base frames for the pipette tips on the system.
14
To resume processing: Turn on the AutoMate using the switch under the monitor.. Wait
for the sequence of three beeps.
15
Press Start
16
Remove the tip carrier from the cleaning solution after 30 minutes.
17
Dry the exterior with a lint-free absorbent tissue.
18
Clean inside the tip carrier duct using a cleaning brush (B37552).