Run the System
Detailed Rack Information
Placing unfinished MTPs in manual mode
When you place unfinished MTPs for a second run in manual mode, follow these steps:
1
Use the same MTP IDs. Enter the IDs and start position.
2
Return the MTP to the same position used.
MTP Display
The MTP module status is displayed on screen. Display options described can differ
depending on your MTP configuration.
After drawer initialization, the background color changes to orange.
Figure 2.22 MTPs during initialization
If the MTP drawer displays a red X, communication is being tested with the LIS. The LIS
queries the following information:
• Are the MTP bar code labels known?
— The LIS only accepts MTPs that it recognizes.
2-60
Caution
Risk of sample mismatch. When using unfinished plates a second time, place the
plates in same position previously used to prevent a sample mismatch in the
system.
B67471B